Change 7

This function is available when the volume label on the C-drive is IZ202 or higher.
 User recovery function

Change 6

"System Recovery" can not be applied. Reffer to the following explation.
System recovery
NOTE
The 4287A with option 011 can not operate this function.
Performing system recovery will allow you to reset the Windows operating system and the
firmware of the 4287A to the state when you purchased
If, for some reason, the Windows operating system or firmware failed and cannot be started
normally or operation after startup is unstable, execute system recovery.
Notes on executing system recovery
Performing system recovery causes the following:
 The following settings of the 4287A are initialized.
 If the firmware has been updated after purchasing the 4287A, the firmware when you
purchased the product
Files you created using the save function are not affected, but we recommend backing them
up before executing system recovery for precautionary purposes.
Procedure to execute system recovery
NOTE
You need the keyboard for this operation.
Step 1. Shut down the 4287A.
Step 2. Connect the keyboard to the 4287A.
Step 3. Press the standby switch of the 4287A to turn it on.
Step 4. When the message in the figure below appears on the screen following the Agilent's logo
Appendix A
Network settings
GPIB settings
Printer settings
*1
is recovered.
*1.If the hard disk failed and has been replaced after purchase, the state when the
replacement was performed is recovered.
Manual Changes
Manual Changes