5. Final test

5.1 General

Important!
A final test is required after each and every repair.
Please maintain a record of all maintenance and
service work that you perform. We suggest that you
maintain this record on a service record form, as
found on page 34 of this manual. You may photocopy
this page for manual entry, or obtain the manual in
PDF format, display it using Acrobat Reader, complete
it on a PC, and print it.
Enter the following values for SOMNOcomfort 2/
SOMNOcomfort 2e in your service record (see
page 42):
Operating hours
Humidity level
Softstart time and initial pressure
Patient's therapy pressure
Enter the following values for SOMNObalance/
SOMNObalance e in your service record (see
page 42):

5.2 Performing the test

5.2.1 Preparation

Assemble the therapy unit ready to use
including breathing tube
supply
and power supply cable
3

5.2.2 Checking the power supply cable

1. Check the power supply cable
Make sure that
– insulation is intact,
– the cable is not damaged,
– and that there are no loose connections.
2. Replace the power supply cable
necessary.
, mask
, power
9
11
.
2
.
2
if
2
Operating hours
Humidity level
Softstart time and initial pressure
Mode
for CPAP: therapy pressure
for APAP: bottom and top pressure limit, rise
speed as per patient record
If you detect errors or deviations from defined
values in the course of your final testing you must
not redeploy SOMNOcomfort 2/
SOMNOcomfort 2e/SOMNObalance/
SOMNObalance e before these errors have been
eliminated. You can determine possible causes of
such errors and recommended countermeasures by
referring to chapter "6. Troubleshooting" on
page 19.