Removing The Diagnostic Cell - Thermo Scientific Vanquish VH-D10 Operating Manual

Diode array detectors
Table of Contents

Advertisement

Follow these steps
7.8.2

Removing the Diagnostic Cell

Preparations
Follow these steps
Diode Array Detectors (VH-D10)
Operating Manual
1. Remove the shipping locks carefully on the left and right rear side of
the diagnostic cell.
TIP Store the shipping locks in the cell packaging to have them easily
available when storing or shipping the cell.
2. Check the position of the rotating locks on the front of the
diagnostic cell.
If the rotating locks are not in a horizontal position, turn them
counterclockwise to a horizontal position. To insert the cell, the
rotating locks must always be in a horizontal position and thus
opened.
3. Carefully insert the diagnostic cell into the flow cell opening of the
device. The diagnostic cell must sit completely in the device.
4. Turn the rotating locks simultaneously clockwise until they are in a
vertical position. The diagnostic cell is locked in place, when the
rotating locks arrest.
5. Close the device doors to allow detection of the diagnostic cell.
During this, the device reads the information on the ID chip of the
diagnostic cell.
6. If the lamp is turned on, you can open the shutter in the light path
before the diagnostic cell.
If no flow cell is to be installed afterward
Turn off the UV lamp.
If a flow cell is to be installed afterward
Close the shutter in the light path before the diagnostic cell.
1. Turn the rotating locks on the diagnostic cell simultaneously
counterclockwise to a horizontal position to unlock the diagnostic
cell.
2. Carefully pull out the cell from the opening in the device.
3. Store the diagnostic cell in its packaging.
7
Maintenance and Service
Page 143