Manifold Door Hinge - Thermo Scientific TSQ 8000 Evo Hardware Manual

Hide thumbs Also See for TSQ 8000 Evo:
Table of Contents

Advertisement

Manifold Door Hinge

Thermo Scientific
To replace the manifold door hinge:
Note See the TSQ 8000 Evo Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
System"
on
page
107.
Note You do not have to remove the ion source cartridge or the column.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Remove the four screws on the left hand front panel and remove the panel. One is located
at the bottom left side of the front of the instrument.
Figure 223. Locating the Left Hand Front Panel Screws
3
Replacing Components of the Manifold
"Shutting Down the TSQ 8000 Evo
Nuts
Screw Located on Front of
Instrument
TSQ 8000 Evo Hardware Manual
Advanced Troubleshooting
213