14-6
4. To exit any menu without saving parameter changes, press and release the
About UserAtten Mode Parameters
UserAtten
In
update mode (
adjusted using the base attenuation (
parameters. Both base attenuation and reference can be used simultaneously,
producing an actual attenuation equal to (
Manual Operation " on page 14-4, for more information.
BaseAtt‐‐Base Attenuation
Adds a fixed attenuation value, shifting the scale down and allowing attenuation to be
displayed relative to this base level (useful for calibrating signals played over varying
transducers). See "Setting Base Attenuation" on page 14-8, for more information.
StpSize‐‐Step Size
Sets the increments in which attenuation is applied to the signal when using the
Select knob.
Refrnce—Reference
Sets a reference value used to "flip" the scale of the display (useful for displaying
actual signal level on the front panel of the PA5). May be used only when the
intensity of the input signal is known. See "Setting a Reference Value " on
page 14-10, for more information.
Update—Update
Determines whether attenuation changes dynamically as the selector knob is turned or
only after pressing enter to select the current value.
AbsMin‐‐Minimum Attenuation
Sets the minimum level of attenuation the user can apply to the signal (to avoid
accidentally presenting excessively loud signals).
PA5 Manual Operation Menus
To access a menu:
•
For a definition of each menu item:
•
To exit a menu without changing settings:
•
PA5 Programmable Attenuator
ESC button before the settings are saved.
Mode, the user may set parameters such as step size (
Update
), and minimum attenuation (
Turn the knob until the name of the desired menu appears on the display,
then press and release the knob. The module has two levels of menus.
Top-level menu items are indicated by a single filled box in the upper left
corner of the menu display, and sub-menus are indicated with an additional
indicator box for each level. Only the UserOps menu item has sub-menu
items.
Turn the Select knob until the name of the menu appears on the display,
then press and hold down the Select knob. A description of the menu func-
tion will scroll across the display.
Press and release the ESC button.
AbsMin
). The scale can be
BaseAtt
) and reference value (
Refrnce
BaseAtt
+
-dial setting). See "PA5
System 3
StpSize
),
Refrnce
)