TDT RZ2 Manual page 360

System 3 series. rz z series
Table of Contents

Advertisement

7-46
Integer Value" on page 7-44. for more information.
The table below maps the output channels of the RZ5 and RX7 to their poke
address.
Isolator Output
Channel
1
2
3
4
5
6
7
8
Global Reference Enable
Global reference uses the analog ground to complete the stimulation circuit. The
global reference feature can be enabled by setting the value of a specific memory
address to one. The StimRef indicator light on the front panel of the stimulus isolator
is illuminated when the global reference has been set.
RZ5
RX7
Channel Masks
Memory addresses for stimulus, reference, or digital I/O channel setup expect an
integer value between zero and 65535. Masked values for each channel are noted
in the table below. Adding masked values together will set multiple channels.
MS4/MS16 Stimulus Isolator
Poke Waveform To
Address
RZ5
RX7
32
20
9
33
21
10
34
22
11
35
23
12
36
24
13
37
25
14
38
26
15
39
27
16
To enable global reference when using an RZ5 set the value of
memory address 50 to one as pictured above.
To enable global reference when using the RX7 set the value
of address 9 to one.
Poke Waveform To
Isolator Output
Channel
RZ
40
41
42
43
44
45
46
47
System 3
Address
RX7
28
29
30
31
32
33
34
35