Trigger.timer[N].Delay - Keithley 2600B Reference Manual

System sourcemeter instrument
Hide thumbs Also See for 2600B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual

trigger.timer[N].delay

This attribute sets and reads the timer delay.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
interval = trigger.timer[N].delay
trigger.timer[N].delay = interval
interval
N
Details
Once the timer is enabled, each time the timer is triggered, it uses this delay period.
Assigning a value to this attribute is equivalent to:
trigger.timer[N].delaylist = {interval}
This creates a delay list of one value.
Reading this attribute returns the delay interval that will be used the next time the timer is triggered.
Example
trigger.timer[1].delay = 50e-6
Also see
trigger.timer[N].reset()
trigger.timer[N].delaylist
This attribute sets an array of timer intervals.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
intervals = trigger.timer[N].delaylist
trigger.timer[N].delaylist = intervals
intervals
N
2600BS-901-01 Rev. C / August 2016
Affected by
Instrument reset
Recall setup
Trigger timer N reset
Delay interval in seconds (5.00e-07 to 100,000)
Trigger timer number (1 to 8)
(on page 7-378)
Affected by
Instrument reset
Recall setup
Trigger timer N reset
Table of delay intervals in seconds
Trigger timer number (1 to 8)
Section 7: TSP command reference
Where saved
Default value
Not saved
10e-6 (10 µs)
Set the trigger timer 1 to delay for
50 µs.
Where saved
Default value
Not saved
{10e-6}
7-375

Advertisement

Table of Contents
loading

Table of Contents