Table 4-4 Status Byte Register
Bit
Description
0
0: Measurement data is not ready in the output queue.
1: Measurement data is ready in the output queue.
1
0: Measurement is completed.
1: Measurement is not completed.
2, 3
These bits are always set to 0.
4
0: Measurement data is not ready in the output queue.
1: Measurement data is ready in the output queue.
5
0: Standard Event Status Register Masked by Standard Event Status Enable Register is not set to 1.
1: Standard Event Status Register Masked by Standard Event Status Enable Register is set to 1.
6, 7
These bits are always set to 0.
N9360A Multi UE Tester CDMA2000 Programming Manual
Status Byte Register
The Status Byte Register of the Tester uses the bit 0
Measurement Data Ready Bit, bit 1 Measuring Status Bit, bit 4
Message Available and bit 5 Standard Event Status Summary Bit as
shown in
Figure
4-7.
Figure 4-7 Status Byte Register
Bit Number
7
6
The Status Byte Register contains the following information:
To query the Status Byte Register, send the command *STB?.
The response is the decimal sum of the bits which are set to
For example, if the bit 5 and bit 4 are set to
of these 2 bits is 48 (32 plus 16) see
value 48 is returned.
Programming Command Guidelines
5
4
3
2
1
*
STB?
Figure
4
0
.
1
, the decimal sum
1
4-5. So the decimal