Fixture Compensation; Performing Fixture Compensation - Agilent Technologies 42941A Operation And Service Manual

Impedance probe kit
Table of Contents

Advertisement

Operation

Fixture Compensation

Fixture Compensation
In an actual measurement, a probe adapter is attached to the probe. Fixture
compensation is required for compensating residual impedance and admittance.
The fixture compensation includes OPEN and SHORT compensation
measurements. For more information on fixture compensation, also refer to the
4294A Operation Manual.
NOTE
Generally, there is no need to perform load compensation. However, if you have a
suitable standard device and require extremely accurate measurement, perform
load compensation as needed.
The following procedure is for measurement of the compensation data with the
42941A.

Performing Fixture Compensation

Attach the probe adapter to be used for your measurement and perform fixture
compensation.
1.
2.
Performing Open Compensation
1. Attach the probe adapter to the probe but leave it disconnected from a device.
Figure 3-5
Performing Open Compensation
22
Press [Cal] key to bring up the Calibration Menu.
Press FIXTURE COMPEN key to bring up the Fixture Compensation Menu.
Chapter 3