Connecting the Test Head

When you connect or replace the test head, turn o the 4286A.
1.
Place the test xture stand on a level surface. At this time, make sure that the
hexagonal hole on top of the stand is located in front of the two mounting posts as you look
at the stand. Then, lay the stand on its side.
2.
Allow the test head to pass under the xture stand from the side opposite to yours.
3.
Position the test head such that the connectors of the test head and holes on top of the
xture stand are aligned.
4.
Slightly tighten the two screws on the bottom surface of the xture stand to ensure
that the test head is suciently stable. Press screws to check for correct engagement of
their thread.
5.
Further tighten the screws while holding the test head with your hand to completely
secure the head in place.
6.
Place the xture stand with its bottom surface down.
Connecting the Test Head with L-type Coaxial Adapter
(Option 022 only)
When you connect or replace the test head, turn o the 4286A.
1.
Connect Connecting L-type Coaxial Adapter to the Test Head.
2.
Place the test xture stand on a level surface. At this time, make sure that the
hexagonal hole on top of the stand is located in front of the two mounting posts as you look
at the stand. Then, lay the stand on its side.
3.
Allow the test head to pass under the xture stand from the side opposite to yours.
4.
Position the test head such that the connectors of the L-type Coaxial adapter and holes
on top of the xture stand are aligned.
5.
Slightly tighten the two screws on the bottom surface of the xture stand to ensure
that the test head is suciently stable. Press screws to check for correct engagement of
their thread.
Installation and Set Up Guide
2-9