Y Scale; Auto Scale - Agilent Technologies 86038B User Manual

Photonic dispersion and loss analyzer
Table of Contents

Advertisement

Y Scale

Auto Scale

120
Ref Level
Specifies the reference level value, the maximum expected value
to be measure, at the reference level position. The Y-scale units
depend on the type of measurement.
• DGD and GD measurements are specified in ps.
• Gain/Loss and PDL measurements are specified in dB.
• CD measurements are specified in ps/nm or ps/nm-km.
• PCD is specified in ps/nm.
• Depolarization Rate is specified in ps.
• 2nd order PMD is specified in ps
• Optical phase is specified in radians
Units/Division
Specifies the units per division of the vertical scale.
Ref. Level Position
Sets the position of the reference level to the chosen scale line
on the graph, where 0 is the bottom and 10 is the top line.
View Full Range
Ensures that the highest and lowest Y-value is visible on the
trace.
Snap Enabled
When selected, the mouse pan and zoom functions are snapped
to a rounded value of amplitude units. This is useful when you
want to keep the units in an easy to calculate form. Clear Snap
Enable when you need precision tuning of a portion of the graph
where the exact settings are important.
Centers the graph around the response. This function is useful
after using the zoom functions.
Agilent 86038B Photonic Dispersion and Loss Analyzer, Second Edition
2
.