Selecting Measurement Parameters; Definition Of S-Parameters; Setting Up S-Parameters - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Setting Measurement Conditions

Selecting Measurement Parameters

Selecting Measurement Parameters
The E5070B/E5071B allows users to evaluate the DUT (device under test) characteristics
by using the following measurement parameters.
S-parameters
Mixed mode S-parameters
This section gives the definition of S-parameters and explains how to choose their values.
For the definition and use of mixed mode S-parameters, refer to "Evaluating Balanced
Devices (balance-unbalance conversion function)" on page 293.

Definition of S-parameters

S-parameters (scattering parameters) are used to evaluate how signals are reflected by and
transferred through the DUT. An S-parameter is defined by the ratio of two complex
numbers and contains information on the magnitude and phase of the signal. S-parameters
are typically expressed as follows.
S
out in
out: port number of the DUT from which the signal is output
in: port number of the DUT to which the signal is input
For example, S-parameter S
input signal of port 1 on the DUT, both expressed in complex numbers.

Setting up S-parameters

Step 1. Press
which measurement parameters will be set up.
Step 2. Press
.
Step 3. Press a softkey that corresponds to the desired S-parameter.
S-parameters on the softkeys are expressed as follows.
S
out in
out: test port number of the E5070B/E5071B to which the DUT's output signal is input
in: test port number of the E5070B/E5071B from which the signal is applied to the DUT
80
is the ratio of the output signal of port 2 on the DUT with the
21
(or
) and
(or
) to select the trace for
Chapter 3