Set Up the Test Equipment

16962A Logic Analyzer Service Guide
This section explains how to set up the test equipment for
the minimum master- to- master clock time/maximum state
data rate test.
Turn on the required test equipment. Let all of the test
1
equipment and the logic analyzer warm up for 30 minutes
before beginning any test.
Set up the pulse generator according to one of the
2
following tables.
Set the frequency of the pulse generator:
a
In this test procedure, the logic analyzer uses both
edges of the clock to acquire data. The test frequency
is half the test clock rate because data is acquired on
both the rising edge and the falling edge of the clock.
Set the frequency to 1000 MHz plus 2% (1020 MHz).
This includes the frequency uncertainty of the pulse
generator, plus a test margin.
For example, if you are using an 8133A pulse
generator, the frequency accuracy is ±0.5% of setting.
If you are using an 81134A pulse generator, the
frequency accuracy is ±0.005% of setting.
Set the rest of the pulse generator parameters to the
b
values shown in one of the following tables.
Testing Logic Analyzer Performance