Measuring Status Register Set; Averaging Status Register Set - Agilent Technologies 8712ET Programmer's Manual

Rf network analyzers
Hide thumbs Also See for 8712ET:
Table of Contents

Advertisement

Measuring Status Register Set

The Measuring Status register set monitors conditions in the analyzer's
measurement process.
Bits in the Measuring Status condition register are set to 1 under the
following conditions:
Channel 1
Measuring
(bit 0) is set to 1 while the analyzer is collecting
measurement data on channel 1.
Channel 2
Measuring
(bit 1) is set to 1 while the analyzer is collecting
measurement data on channel 2.

Averaging Status Register Set

The Averaging Status register set monitors conditions in the analyzer's
measurement process when the trace averaging function is in use.
Bits in the Averaging Status condition register are set to 1 under the
following conditions:
Measurement
Channel 1
Averaging
(bit 0) is set to 1 while the analyzer is sweeping on
measurement channel 1 and the number of sweeps
completed (since "average restart") is less than the
averaging factor.
Measurement
Channel 2
Averaging
(bit 1) is set to 1 while the analyzer is sweeping on
measurement channel 2 and the number of sweeps
completed (since "average restart") is less than the
averaging factor.
Programmer's Guide
Using Status Registers
The Analyzer's Status Register Sets
5-23