Average-only Mode
• Absolute:Logarithmic 0.02 dB Linear 0.5%
• Relative:Logarithmic 0.04 dB Linear 1.0%
Normal Mode
• Absolute and relative accuracy vs. temperature is shown in the following
table:
Absolute accuracy (log)
Absolute accuracy (linear)
Relative accuracy (log)
Relative accuracy (linear)
1. Power meter is within 5 ºC of its calibration temperature.
Time Base
0.01%.
Zero Set, Zero Drift and Measurement Noise:
For E- Series E9320 sensors, refer to the Specifications Section in the E- Series
E9320 Power Sensors Operating and Service Guide.
6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1
Calibration Temperature
Temperature 0 to 55 ºC
±5 ºC
±0.04 dB
±0.08 dB
±0.8%
±1.7%
±0.08 dB
±0.16 dB
±1.6%
±3.4%
E4416A/E4417A Power Meters Service Guide