Reading Only The Actual Measurement Point/Reading The Value Interpolated Between Measurement Points - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Reading only the actual measurement point/Reading the value
interpolated between measurement points
The point on the trace on which a marker can be placed differs depending on how the
discrete marker mode is set up.
Turning on discrete
mode
(
Discrete ON
Turning off discrete
mode
(
Discrete OFF
Figure 6-2
Marker discrete mode
Turning Discrete Mode On or Off
Step 1. Press
which the discrete mode is set up.
Step 2. Press
Step 3. Press
Chapter 6
A marker moves only between actual measurement points. When a
specific marker stimulus value is specified as a numerical value, the
marker is placed at the measurement point closest to the specified
)
value. A marker that is placed between interpolated points with the
discrete mode off automatically moves to the nearest measurement
point when the discrete mode is turned on.
The marker can move from one actual measurement point to
another. Because it is interpolated, it can also move in the space
between measurement points.
)
(or
.
to turn the discrete mode on or off.
Discrete
Analyzing Data on the Trace Using the Marker
) and
(or
Data Analysis
) to activate the trace on
257