Smux.trigger.source.action - Keithley 2657A Reference Manual

High power system sourcemeter instrument
Hide thumbs Also See for 2657A:
Table of Contents

Advertisement

Section 7: TSP command reference

smuX.trigger.source.action

This attribute enables or disables sweeping the source (on or off).
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
action = smuX.trigger.source.action
smuX.trigger.source.action = action
action
X
Details
This attribute is used to enable or disable source level changes during a sweep. In addition to enabling the action
before initiating the sweep, make sure to configure it using smuX.trigger.source.linearY(),
smuX.trigger.source.listY(), or smuX.trigger.source.logY().
Example
smua.trigger.source.listv({3, 1, 4, 5, 2})
smua.trigger.source.action = smua.ENABLE
Also see
smuX.trigger.source.linearY()
smuX.trigger.source.listY()
smuX.trigger.source.logY()
Triggering
(on page 3-31)
7-240
Model 2657A High Power System SourceMeter® Instrument Reference Manual
Affected by
SMU reset
Instrument reset
Recall setup
Sweep source action. Set to one of the following values:
0 or smuX.DISABLE: Do not sweep the source
1 or smuX.ENABLE: Sweep the source
Source-measure unit (SMU) channel (for example,
smua.trigger.source.action applies to SMU channel A)
(on page 7-242)
(on page 7-243)
(on page 7-244)
Where saved
Default value
Not saved
0 (smuX.DISABLE)
Configure list sweep for SMU
channel A (sweep through
3 V, 1 V, 4 V, 5 V, and 2 V).
Enable the source action.
2657A-901-01 Rev. B/December 2012

Advertisement

Table of Contents
loading

Table of Contents