Smux.trigger.measure.action - Keithley 2600B Reference Manual

System sourcemeter instrument
Hide thumbs Also See for 2600B:
Table of Contents

Advertisement

Section 7: TSP command reference
Also see
bufferVar.clearcache()
Configuring and running sweeps

smuX.trigger.measure.action

smuX.trigger.measure.Y()
smuX.trigger.source.action
smuX.trigger.source.linearY()
smuX.trigger.source.listY()
smuX.trigger.source.logY()
Triggering
(on page 3-32)
smuX.trigger.measure.action
This attribute controls measurement actions during a sweep.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
action = smuX.trigger.measure.action
smuX.trigger.measure.action = action
action
X
Details
With this attribute enabled (setting action to smuX.ENABLE or smuX.ASYNC), configure the
measurement with one of the smuX.trigger.measure.Y() functions.
If this attribute is set to smuX.ASYNC:
Asynchronous sweep measurements can only be used with measure autoranging turned off. To turn
measure autoranging off for all measurements during the sweep, set the
smuX.measure.autorangeY attribute to smuX.AUTORANGE_OFF.
Autozero must also be turned off. To turn off autozero, set the smuX.measure.autozero attribute to
smuX.AUTOZERO_OFF or smuX.AUTOZERO_ONCE.
If either of the above items is incorrectly configured, the smuX.trigger.initiate() function
generates an error.
Also see
smuX.trigger.autoclear
smuX.trigger.measure.Y()
Triggering
(on page 3-32)
7-260
Series 2600B System SourceMeter® Instrument Reference Manual
(on page 7-20)
(on page 3-29)
(on page 7-260)
(on page 7-264)
(on page 7-266)
(on page 7-268)
(on page 7-269)
(on page 7-270)
Affected by
SMU reset
Instrument reset
Recall setup
The sweep measurement action; set to one of the following values:
0 or smuX.DISABLE: Do not make measurements during the sweep
1 or smuX.ENABLE: Make measurements during the sweep
2 or smuX.ASYNC: Make measurements during the sweep, but asynchronously
with the source area of the trigger model
Source-measure unit (SMU) channel (for example,
smua.trigger.measure.action applies to SMU channel A)
(on page 7-253)
(on page 7-264)
Where saved
Default value
Not saved
0 (smuX.DISABLE)
2600BS-901-01 Rev. C / August 2016