Preparing Administration Set (
For instructions on how to go from checking in a Pump
Module to preparing it for an infusion setup, reference
"General Setup and Operation".
1.
If a new set is being loaded, prime set (reference
"Priming" procedure).
2.
Open Pump Module door.
3.
Load administration set, as follows:
a. Hold upper fitment above fitment
recess and lower into recess.
b. Ensure tubing is not twisted.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Loading
)
Pump Module
WARNINGS
• To prevent a potential free-flow
condition, ensure no extraneous
object (for example, bedding,
tubing, glove) is enclosed or
caught in the Pump Module door.
• Administration Sets:
Use only Alaris
♦
Module/Gemini Infusion
System administration sets.
The use of any other set may
cause improper instrument
operation, resulting in an
inaccurate fluid delivery or
other potential hazard. For a
list of compatible sets,
reference the Set Compatibility
Card (provided separately).
Discard if packaging is not
♦
intact or protector caps are
unattached.
CAUTIONS
• Before operating instrument,
verify that administration set is
free from kinks and installed
correctly in instrument.
• Insert upper fitment before
installing safety clamp fitment.
• When reloading an
administration set, leave the
safety clamp fitment in the closed
position (reference "General
Information", "Safety Clamp
Fitment").
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pump
®
2-3
Getting Started