Preparing Administration Set (Pump Module) - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Preparing Administration Set (
For instructions on how to go from checking in a Pump
Module to preparing it for an infusion setup, reference
"General Setup and Operation".
1.
If a new set is being loaded, prime set (reference
"Priming" procedure).
2.
Open Pump Module door.
3.
Load administration set, as follows:
a. Hold upper fitment above fitment
recess and lower into recess.
b. Ensure tubing is not twisted.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Loading
)
Pump Module
WARNINGS
• To prevent a potential free-flow
condition, ensure no extraneous
object (for example, bedding,
tubing, glove) is enclosed or
caught in the Pump Module door.
• Administration Sets:
Use only Alaris
Module/Gemini Infusion
System administration sets.
The use of any other set may
cause improper instrument
operation, resulting in an
inaccurate fluid delivery or
other potential hazard. For a
list of compatible sets,
reference the Set Compatibility
Card (provided separately).
Discard if packaging is not
intact or protector caps are
unattached.
CAUTIONS
• Before operating instrument,
verify that administration set is
free from kinks and installed
correctly in instrument.
• Insert upper fitment before
installing safety clamp fitment.
• When reloading an
administration set, leave the
safety clamp fitment in the closed
position (reference "General
Information", "Safety Clamp
Fitment").
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Pump
®
2-3
Getting Started