Preparing Syringe and Administration Set (
Priming (continued)
Pressure Sensing Disc Installed (Continued)
4.
Prime, as follows:
a. Orient pressure sensing disc with patient side up.
b. Depress and hold pressure sensing disc between
fingers.
2
c. Press and hold
priming of syringe administration set is complete.
d. Release pressure sensing disc.
5.
When priming is complete, release
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
soft key until fluid flows and
PRIME
PRIME
Syringe Module
Patient Side
soft key.
FU
RATE (mL/h)
)
(Continued)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
A
Prime Set with Syringe
CAUTION: Do
not prime with
disposable set
connected to
patient!
Prime Volume
0
mL
>Press and Hold PRIME
EXIT
PRIME
A
Prime Set with Syringe
CAUTION: Do
not prime with
disposable set
connected to
patient!
Prime Volume
2 mL
>Press and Hold PRIME
EXIT
PRIME
Getting Started
2-13