8210 Series (Continued)
Alarm
Check Sensor - Weak Pulse
Check Sensor - Weak Signal
8220 Series
Alarm
Check Sensor - Light
Check Sensor - Low Perfusion
Check Sensor - Low Signal I.Q.
8210 Series
Message
Check Sensor - Electrical or
Optical Interference
Check Sensor - High Pulse
Amplitude
Check Sensor - Excessive Ambient
Light
Alaris
System Directions for Use
®
SpO
Modules Section
2
Alarms and Messages (Continued)
Alarms (Continued)
Patient's low perfusion has
inhibited monitoring.
Low quality of signal being
measured.
Light interference on sensor.
Patient's low perfusion has
inhibited monitoring.
Low signal quality being measured.

Messages

External interference on sensor.
Artifact interfering with pulse
reading.
Light interference on sensor.
(8210)
Meaning
(8220)
Meaning
(8210)
Meaning
Troubleshooting and Maintenance
Response
Check sensor. Move sensor to a
better perfused site.
Check sensor. Confirm correct
sensor placement. Move sensor to
a better perfused site.
Response
Check sensor. Remove or reduce
lighting. Cover or reposition
sensor.
Check sensor. Move sensor to a
better perfused site.
Check sensor. Confirm correct
sensor placement. Move sensor to
a better perfused site.
Response
Check sensor. Identify source of
external interference if other than
sensor.
Check sensor. Relocate sensor to
a site with less artifact interference.
Check sensor. Remove or reduce
lighting. Cover or reposition
sensor.