Setting Up Pca Dose Only - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Setting Up PCA Dose Only

1.
Perform steps in "Preparing Infusion".
2.
Press PCA Dose Only soft key from Infusion Mode
screen.
3.
To enter
dose, use numeric data entry keys.
PCA
4.
To enter lockout interval, press
key and use numeric data entry keys.
5.
To enter maximum limit, press
then Yes soft key.
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
Infusion Modes (Continued)
LOCKOUT INTERVAL
MAX LIMIT
soft
soft key and
A
Guardrails Drug Setup
Morphine
INFUSION MODES
PCA Dose only
Continuous Infusion
PCA
Dose + Continuous
Loading Dose Only
>Select an Option
DRUG
SETUP
LIBRARY
A
PCA Only
Morphine
_ 20
PCA
mg
DOSE
LOCKOUT
_ _ _
INTERVAL
minutes
MAX
None
LIMIT
LOADING
None
DOSE
[Conc]: 1 mg/mL
>Select LOCKOUT
INTERVAL
CHANGE
MODE
A
Guardrails Drug Setup
IMS Pre-fill 30mL
PCA
20
mg
DOSE
LOCKOUT
10
minutes
INTERVAL
Yes
MAX
> > > >
No
LIMIT
LOADING
None
DOSE
[Conc]: 1 mg/mL
>Select Max Limit Option
CHANGE
MODE
Programming
3-13