Messages - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Pump and Syringe Modules
Message
Anesthesia Mode
Bolus Dose Complete
Delay Complete
Infusion Complete
Infusion Complete - KVO
Panel Locked
Panel Unlocked
Pause
Start time for next dose has
passed.
2-116
Troubleshooting and Maintenance
Alarms, Errors, Messages (Continued)

Messages

Anesthesia Mode discontinued
when disconnected from
Module running in continuous
infusion mode if programmed.
Delay time completed.
Current infusion completed.
Programmed volume-to-be-infused
delivered; module running at
rate.
Tamper Resist feature is active and
a key was pressed.
Tamper Resist feature deactivated.
Pause control pressed; infusion
stopped.
Start of next dose passed.
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Meaning
.
AC
KVO
Response
Press
soft key.
CONFIRM
None
Press
key, or press
RESTART
key and then
CHANNEL SELECT
soft key.
START
Set up a new infusion or press
key.
CHANNEL OFF
Set up a new infusion or press
key.
CHANNEL OFF
If appropriate, deactivate Tamper
Resist feature using Tamper Resist
Control on back of PC Unit.
None.
To resume infusion, press
key, or press
RESTART
key and then
SELECT
key.
Press
soft key.
CONFIRM
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
CHANNEL
soft
START

Advertisement

Table of Contents
loading