Setting Bolus Dose (Continued)
3.
Set key to
PROGRAM
authorization code and press
4.
To enter dose value, use numeric data entry keys.
5.
Press
CONFIRM
• If programmed bolus dose is outside soft limit for that
care area, a prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
• If programmed bolus dose is outside hard limit for that
care area, a prompt appears before programming can
continue. Infusion needs to be reprogrammed.
6.
If Authorization Code is disabled, door must be locked
prior to starting bolus dose.
7.
Verify dose value is correct and then press
key:
Delivering Bolus
• Bolus and drug name alternate with VTBI in Main
Display
• When bolus dose is complete,
scrolls in Channel Message Display.
• Programmed infusion resumes.
NOTE:
The
BOLUS DOSE
has begun in
PCA
continuous infusion modes.
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
Infusion Modes (Continued)
position or enter
CONFIRM
soft key.
scrolls in Channel Message Display
BOLUS COMPLETE
soft key is only available once an infusion
dose only, continuous infusion, or
-digit
4
soft key.
soft
START
+
PCA
A
Guardrails Drug Setup
IMS Pre-fill 30mL
2
Cont. Dose:
mg/h
30
Max Limit:
mg/4 h
Bolus
1
mg
Dose:
[Conc]: 1 mg/mL
>Press START
PRO-
BOLUS
PAUSE
START
GRAM
DOSE
Programming
3-21