Primary Infusion
Bolus Dose (Continued)
Stopping Bolus Dose (Continued)
3.
To stop bolus and start continuous infusion, press Yes
soft key.
4.
To stop continuous infusion, press and hold
key until a beep is heard (approximately
OFF
seconds).
1.5
Restoring Bolus Dose
A bolus dose can be restored after it has completed, either
prior to or after the module has been turned off, as indicated in
the following procedures.
The display examples in this procedure represent restoring a
bolus dose which was programmed using the Drug Library.
Even where the displays are different when restoring a bolus
dose which was programmed using a non-library drug, the
procedure is the same.
Bolus dose completed - module not turned off:
1.
a. Press
CHANNEL SELECT
b. Verify infusion parameters and press
2-30
Programming
- With Guardrails
Suite MX Protection (Continued)
®
key.
BOLUS
CHANNEL
soft key.
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Setup
A
Alteplase
Stop Bolus and Start
Yes
Continuous infusion?
No
>Press Yes or No
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Alaris
System Directions for Use
®
Pump and Syringe Modules Section