Preparing Administration Set (
6.
To prime tubing and clear air from injection sites and
tubing fitments, slowly open roller clamp.
7.
When priming is complete, close roller clamp.
8.
Verify no fluid flow.
Preparing Syringe and Administration Set (
For instructions on how to go from checking in a Syringe
Module to preparing it for an infusion setup, reference
"General Setup and Operation".
1.
Prepare syringe (reference "General Information",
"Compatible Syringes") in accordance with
manufacturer's directions for use.
2.
Prepare administration set (reference Set Compatibility
Card, provided separately) in accordance with
manufacturer's directions for use.
3.
Attach upper fitting of administration set to syringe tip.
2-6
Getting Started
Priming (Continued)
) (Continued)
Pump Module
Syringe Module
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
FU
RATE (mL/h)
)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
WARNING
Use only standard, single-use,
disposable syringes (with luer-lock
connectors) and administration sets,
designed for use on syringe pumps.
The use of any other syringe or
administration set may cause
improper instrument operation,
resulting in inaccurate fluid delivery
or pressure sensing, or other
potential hazards. For a list of
compatible syringes, reference
"General Information", "Compatible
Syringes". For a list of compatible
administration sets, reference the
Set Compatibility Card (provided
separately).
Alaris
System Directions for Use
®
Pump and Syringe Modules Section