Cardinal Health Alaris 8000 Series Directions For Use Manual page 312

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Microstream
Disposables are single-use, disposables which
®
must be changed with each use. The manufacturer's sample
flow,
ml/min, does not affect the disposable's life; however,
50
humidity and specific patient conditions may shorten the
effective life of the disposables. Microstream
are rated for up to
24
the specific Microstream
The
t
Module provides readings in compliance with
E
CO
2
(body temperature, pressure, saturation) standards.
no affect on accuracy due to cyclic pressure up to
NOTE:
(body temperature, pressure, saturation assumed
BTPS
mmHg) calculations are made according to:
47
=
x (Pb -
PCO
FCO
2
2
Where:
is fractional concentration of
FCO
2
=
FCO
% CO
2
Pb is ambient pressure.
is partial pressure of
PCO
2
5-26
General Information
Principle of Operation (Continued)
hours and
hours use, depending on
72
Disposable.
®
)
47
CO
/
.
100
2
at
CO
BTPS
2
Disposables
®
BTPS
There is
kPa.
10
,
37°C
in dry gas and
2
.
Alaris
System Directions for Use
®
EtCO
Module Section
2