Changing Syringe During Infusion (Syringe Module) - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Changing Syringe During Infusion (
1.
To stop infusion, press
2.
Open plunger grippers and syringe barrel clamp.
• An audio prompt sounds (to silence, press
key).
• Red Alarm Status Indicator flashes.
CHECK SYRINGE
3.
Remove syringe and separate administration set from
syringe.
4.
Reattach administration set to new syringe and load new
syringe (reference "Getting Started").
5.
Select syringe type and size (reference "Programming",
"Primary Infusion - With Guardrails
Protection").
6.
Press
CONFIRM
7.
Prime administration set (reference "Getting Started",
"Priming" procedure).
8.
Press
RESTORE
OR
To enter
and rate, press
VTBI
numeric data entry keys, and then
numeric data entry keys.
9.
To begin infusion, press
2-72
General Setup and Operation
System Start-Up / Setup (Continued)
key.
PAUSE
scrolls in Message Display.
Suite MX
®
soft key.
soft key.
soft key and use
RATE
VTBI
soft key.
START
)
Syringe Module
SILENCE
soft key and use
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
SILENCE
OPTIONS
1
2
4
5
7
8
0
CLEAR
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
SYSTEM
ON
3
6
9
ENTER
.
CANCEL