7.

DRIVER SET-UP, OPERATION, AND BASIC INSPECTION

7.1

Mounting Bracket and Clamp Assembly

The Mounting Bracket and Clamp Assembly can be installed in two (2) different orientations.
Both are depicted below. Removing the Mounting Bracket and Clamp Assembly is not
necessary to service the driver.
1.
Place nCPAP driver upside down on a clean, dry surface.
2.
Use a 9/64" Allen wrench to attach the Mounting Bracket Assembly with
the three (3) #8 mounting screws and lock washers included with the
driver. Torque the screws to 7 in-lb.
Mounting Bracket and Clamp Assembly,
Rear Installation-Standard Orientation
7.2

Attach Driver to Stand

Securely attach the driver to a stand or other mounting
system that includes a heater/humidification system. A
Mounting Bracket and Clamp Assembly is included and
should be attached to the driver using the supplied hardware.
Ensure that the stand is rated to safely handle the weight and
dimensions of the driver, as detailed in the "Product
Specification" section, or use the recommended stand as
listed in the approved accessories list.
© © © © 2008, Cardinal Health
AirLife
Mounting Bracket and Clamp Assembly,
Page 21 of 81
®
nCPAP Driver Service Manual
Optional Side Orientation
The Driver Attached to a Stand