Setting Loading Dose Only - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Setting Loading Dose Only

The following procedures should be used when setting a
LOADING DOSE ONLY
Setting Loading Dose
from Infusion Mode Screen
1.
Perform steps in "Preparing Infusion".
2.
Press
LOADING DOSE ONLY
screen.
3.
To enter dose value, use numeric data entry keys.
4.
Verify dose value is correct and then press
key.
• If programmed loading dose is outside soft limit for that
care area, a prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
• If programmed loading dose is outside soft limit for that
care area, a prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
5.
Close and lock security door.
6.
Verify parameters on summary screen are correct and
press
soft key.
START
DELIVERING LOAD
Display.
• Infusion mode and drug name alternate with
Main Display.
• When loading dose is complete, The Loading Dose
has Completed appears on Main Display.
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
Infusion Modes (Continued)
using the Drug Library.
soft key from Infusion Mode
scrolls in Channel Message
soft
CONFIRM
in
VTBI
A
Guardrails Drug Setup
IMS Pre-fill 30mL
_ _ 1
LOADING
mg
DOSE
[Conc]: 1 mg/mL
>Close and Lock Door Then
Press CONFIRM
CHANGE
CONFIRM
MODE
A
Second Nurse Summary
Morphine
Loading
1
mg
Dose:
[Conc]: 1 mg/mL
>Press START
PRO-
PAUSE
START
GRAM
Programming
3-19