Troubleshooting And Maintenance; General; Alarms And Messages; Alarms - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

The
p
Module Technical Service Manual is available from
S
O
2
Cardinal Health. It includes routine service schedules,
interconnect diagrams, component parts lists and
descriptions, test procedures, and other technical information,
to assist qualified service personnel in repair and maintenance
of the instrument's repairable components. Maintenance
procedures are intended to be performed only by qualified
personnel, using the Service Manual, Maintenance Software
and Software User Manual.
Reference the PC Unit Section of this
system references:
Alarms, Errors, Messages
Audio Characteristics
Definitions
Radio Frequency Note
8210 and 8220 Series
Alarm
Bad Sensor
Check Sensor - Electrical or
Optical Interference
High Pulse Rate Alarm
High SpO
Alarm
2
Alaris
System Directions for Use
®
SpO
Modules Section
2

Troubleshooting and Maintenance

General

Alarms and Messages

for the following
DFU

Alarms

Meaning
Broken, unknown or nonsystem
sensor or patient cable attached.
External interference on sensor.
High pulse rate alarm limit has
been exceeded.
High SpO
alarm limit has been
2
exceeded.
Response
Check sensor and patient cable.
Confirm correct sensor and patient
cable are chosen.
Check sensor. Identify source of
external interference if other than
sensor.
Assess patient's condition.
Confirm correct alarm limit values
are selected.
Assess patient's condition.
Confirm correct alarm limit values
are selected.
Troubleshooting and Maintenance
4-35