Delaying Infusion (Continued)
Specifying by Time of Day
The Delay until option is used to program an infusion delay
for a minimum of
minute and up to
1
1.
Press
DELAY OPTIONS
2.
Press Delay until soft key.
3.
If Current time displayed is correct, press
key; otherwise, press Change Time and enter correct
time. (Reference "System Options", "Time of Day" in PC
Unit Section of this
4.
To enter time of day infusion is to be initiated (up to
hours
minutes), use numeric data entry keys.
23
59
5.
Press
CONFIRM
• Time infusion is scheduled to start appears on Main
Display.
• If a Before callback has not been scheduled
(reference "Scheduling a Callback" procedure),
infusion automatically initiates at end of delay period.
NOTE:
If the current time has been previously confirmed, the Time of
Day screen will not be displayed.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Delay Options (Continued)
hours
23
soft key.
.)
DFU
soft key.
minutes.
59
soft
CONFIRM
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A
Delay Options
08:00
Pause
Delay for
Delay until
CALL
BACK
>Select a Delay Option
EXIT
CONFIRM
A
Delayed Options
Time of Day
Current time:
Change
Time
08:00
>CONFIRM Time-of-Day
EXIT
CONFIRM
Programming
2-59