Primary Infusion
Continuous Infusion (Continued)
Changing Rate or VTBI During Infusion
1.
Press
CHANNEL SELECT
2.
Press either
RATE
3.
To enter desired parameter, use up/down arrows for rate
titration, or numeric data entry keys.
4.
Verify correct infusion parameter entry and press
soft key.
A bolus dose can be programmed at the beginning of, or
during, an infusion. The drug being programmed must be a
bolusable drug selected from the Guardrails
non-library drug, as described in the following procedures.
1.
Set up infusion as described in "Continuous Infusion"
procedure, but do not start infusion.
2.
Press
soft key.
BOLUS
• If programmed continuous dose infusion is outside Soft
Limit for that care area, a prompt appears before
programming can continue. If Yes soft key is pressed,
programming continues; if No soft key is pressed,
infusion needs to be reprogrammed.
• If programmed continuous dose infusion is outside
Hard Limit for that care area, a prompt appears before
programming can continue. Infusion needs to be
reprogrammed.
2-26
Programming
- With Guardrails
key.
or
soft key.
VTBI
Bolus Dose
®
Suite MX Protection (Continued)
®
START
Drug Library or a
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Setup
A A A
Alteplase
CONTINUOUS INFUSION
56
RATE
mL/h
100
VTBI
mL
.8
DOSE
mg/kg/h
[Conc]: 1 mg/mL
>Press START
PAUSE
SETUP BOLUS
START
Alaris
System Directions for Use
®
Pump and Syringe Modules Section