Cardinal Health Alaris 8000 Series Directions For Use Manual page 82

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Primary Infusion
Continuous Infusion (Continued)
Changing Rate or VTBI During Infusion
1.
Press
CHANNEL SELECT
2.
Press either
RATE
3.
To enter desired parameter, use up/down arrows for rate
titration, or numeric data entry keys.
4.
Verify correct infusion parameter entry and press
soft key.
A bolus dose can be programmed at the beginning of, or
during, an infusion. The drug being programmed must be a
bolusable drug selected from the Guardrails
non-library drug, as described in the following procedures.
1.
Set up infusion as described in "Continuous Infusion"
procedure, but do not start infusion.
2.
Press
soft key.
BOLUS
• If programmed continuous dose infusion is outside Soft
Limit for that care area, a prompt appears before
programming can continue. If Yes soft key is pressed,
programming continues; if No soft key is pressed,
infusion needs to be reprogrammed.
• If programmed continuous dose infusion is outside
Hard Limit for that care area, a prompt appears before
programming can continue. Infusion needs to be
reprogrammed.
2-26
Programming
- With Guardrails
key.
or
soft key.
VTBI
Bolus Dose
®
Suite MX Protection (Continued)
®
START
Drug Library or a
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Setup
A A A
Alteplase
CONTINUOUS INFUSION
56
RATE
mL/h
100
VTBI
mL
.8
DOSE
mg/kg/h
[Conc]: 1 mg/mL
>Press START
PAUSE
SETUP BOLUS
START
Alaris
System Directions for Use
®
Pump and Syringe Modules Section

Advertisement

Table of Contents
loading

Table of Contents