Secondary Infusion
2.
Press
SECONDARY
3.
Press soft key next to desired drug.
• If applicable, an optional hospital-defined therapy or
clinical indication for delivery of this infusion may
appear. Different limits can be defined for same drug
with different therapeutic indications.
• If applicable, a weight-based, nonweight-based, or
-based option for delivery of this infusion may
BSA
appear.
• If applicable, multiple concentration listings for delivery
of this infusion may appear.
4.
To continue programming, press Yes soft key.
OR
To change selection, press No soft key.
• If Yes was selected and facility has defined a Clinical
Advisory for that drug, a message appears. To
indicate information has been noted and continue
programming, press
• If Yes was selected to continue programming, drug
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
• If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
-- Continued on Next Page --
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
- With Guardrails
(
Pump Module
Infusion (Continued)
soft key.
soft key.
CONFIRM
Suite MX Protection
®
) (Continued)
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Guardrails Fluid Setup
A
0.9% NaCI
PRIMARY INFUSION
40
mL/h
RATE
1000
VTBI
mL
>Press START
DELAY
VOLUME
SECOND-
START
OPTIONS
DURATION
ARY
A
A
Guardrails Drugs
Peds Oncology
Amifostine
A-E
Azathioprine
F-J
Bleomycin
K-O
Cytarabine
P-T
Daunorbicin
U-Z
>Select Drug
PAGE
BASIC
PRIMARY
SEC
DOWN
A
Guardrails Drug Setup
Peds Oncology
Methotrexate
Yes
_ _ _ mg/_ _ _mL was
No
selected.
Is this correct?
2
BSA
DOSING
mg/m
UNITS
>Press Yes or No
Programming
2-41