Powering On System
1.
Connect PC Unit to an external
2.
Press
SYSTEM ON
3.
System self test begins:
• Diagnostics test causes all
Status Indicator lights of attached module(s) to
illuminate briefly.
• Power Indicator illuminates.
• Appropriate module identification (
displays on attached module(s).
• An Audio tone sounds.
• If
Reminder option is enabled and scheduled
PM
preventive maintenance is due,
screen appears.
REMINDER
• At completion of system-on test, New Patient? screen
appears.
NOTES:
Previous infusion parameters are automatically cleared after
hours.
8
If any of the following conditions are observed, the PC Unit or
the affected attached module must be removed from use and
inspected by qualified personnel:
segments are not illuminated during system-on test.
LED
• Indicator lights do not illuminate.
• Appropriate module identification (
displayed.
• Audio tone does not sound.
• Main Display does not appear backlit, appears irregular, or
has evidence of a row of pixels not functioning properly.
If the affected module operates normally when it is attached
via an alternate
replacement module can be substituted.
Alaris
System Directions for Use
®
PC Unit Section

Start-Up

power source.
AC
.
display segments and
LED
MAINTENANCE
,
A
connector, it may be used until a
IUI
,
,
or
)
A
B
C
D
,
or
) is not
B
C
D
General Setup and Operation
1-5