Administration Set / Syringe Information (Continued)
Safety Clamp Fitment (
Safety Clamp Fitment in Closed Position
When an Alaris
Pump Module/Gemini Infusion System
®
administration set is removed from the Pump Module, the
instrument automatically engages the safety clamp fitment in
the closed position (white slide clamp projects out from under
blue fitment). In this closed position, flow is occluded.
Compatible Syringes (
The Syringe Module is calibrated and labeled for use with the
following single-use disposable luer-lock syringes. Use only
the syringe size and type specified on the Main Display. The
full list of permitted syringe models is dependent on the
Syringe Module's software version.
Manufacturer
AstraZeneca
B-D Plastipak
IVAC
Monoject
Terumo
NOTES:
Prefilled Diprivan.
The Monoject SoftPack Luer-Lock Syringe (blister pack) is the
only currently supported Monoject
The Terumo
cc can also be used as a
5
cc.
12
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Pump Module
Syringe Module
1cc
3cc
5cc
cc.
3
) (Continued)
)
6cc
10cc
12cc
cc and the
cc as a
6
10
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Blue Fitment
White Slide Clamp
(shown in closed position)
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
CAUTION
When using a
cc or smaller
10
syringe, Cardinal Health strongly
recommends using an extension set
with a pressure disc, for improved
pressure monitoring and shorter
times to occlusion alarm.
20cc
30cc
35cc
General Information
50cc
60cc