Cardinal Health Alaris 8000 Series Directions For Use Manual page 167

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Pump Module
Alarm
Accumulated Air-in-Line
Air-in-Line
Check IV Set
Close Door
Flo-Stop Open - Close Door
Occluded - Fluid Side/Empty
Container
Occluded - Patient Side
Partial Occlusion - Patient Side
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
Alarms, Errors, Messages (Continued)
Alarms (Continued)
A large number of air bubbles
smaller than current air-in-line limit
has recently passed detector.
Air has been detected in
administration set during an
infusion. Infusion stops on affected
module.
Administration set is not properly
installed. Infusion stops on
affected module.
Door opened during an infusion.
Infusion stops on affected module.
Safety clamp device is in open
position while door is open.
Indicates either upstream occlusion
or empty container. Infusion stops
on affected module.
Increased back pressure sensed
while infusing in pump delivery
mode. Infusion stops on affected
module.
Partial occlusion of patient side of
IV line detected by Auto-Restart
feature.
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
Meaning
Troubleshooting and Maintenance
Response
Clear air from line. To continue
infusion, press
RESET
then
key.
RESTART
Ensure tubing is properly installed
in Air-in-Line Detector. If air is
present, clear air from
administration set. Press
key, or press
CHANNEL SELECT
key and then
soft key.
START
Close roller clamp, remove and
reinstall administration set, close
door, open roller clamp, and then
press
key.
RESTART
Close door. Press
RESTART
or press
CHANNEL SELECT
and then
soft key.
START
Close roller clamp on
administration set or close door.
Clear occlusion on fluid side of
instrument. If necessary, refill drip
chamber. Press
RESTART
press
CHANNEL SELECT
then
soft key.
START
Clear occlusion. Press
key, or press
CHANNEL SELECT
key and then
soft key.
START
Clear occlusion. Press
key, or press
CHANNEL SELECT
key and then
soft key.
START
soft key and
RESTART
key,
key
key
or
,
key and
RESTART
RESTART
2-111