Reviewing or Changing PCA Pause Alarm Limits (Continued)
4.
Verify
pause limits as per facility protocol or
PCA
physician order.
5.
To change
PCA
corresponds to alarm limit and enter a value within
acceptable range.
6.
Press
CONFIRM
7.
Press
soft key.
START
NOTE:
The acceptable range for PCA Pause Protocol is configurable
and defined by the hospital within the data set using the
Guardrails
Suite MX.
®
The
PCA PAUSE LIMITS
. A prompt is provided if the
ALARM LIMITS
must be modified.
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
pause limits, press soft key that
soft key.
must be lower than the
/
SPO2
ETCO2
PCA PAUSE LIMITS
A
PCA Pause Limits
PCA
SPO2/ETCO2
PAUSE LIMITS
ALARM LIMITS
97
88
%SPO2
LOW
5
4
RR
LOW
>Press CONFIRM to Apply
Changes
DISABLE
DISABLE
CONFIRM
SPO2
ETCO2
A
PCA Pause Limits
PCA
SPO2/ETCO2
PAUSE LIMITS
ALARM LIMITS
88
97
%SPO2
LOW
5
_4
RR
LOW
>Enter a Pause Limit
Between 2 - 4
DISABLE
DISABLE
CONFIRM
SPO2
ETCO2
Programming