Primary Infusion
Intermittent Infusion
When using a drug listed in the Guardrails
drug parameters are automatically delivered, based on:
• drug selected
• weight or body surface area (
• dose entry
• rate or duration dose entry
entry
VTBI
1.
Press Guardrails Drugs soft key.
2.
Press soft key next to desired drug.
• If applicable, an optional hospital-defined therapy or
clinical indication for delivery of this infusion may
appear. Different limits can be defined for same drug
with different therapeutic indications.
• If applicable, a weight-based, nonweight-based, or
-based option for delivery of this infusion may
BSA
appear.
• If applicable, multiple concentration listings for delivery
of this infusion may appear.
3.
To continue programming, press Yes soft key.
OR
To change selection, press No soft key.
• If Yes was selected and facility has defined a Clinical
Advisory for that drug, a message appears. To
indicate information has been noted and continue
programming, press
• If Yes was selected to continue programming, drug
amount and diluent volume (if defined in Drug Library)
are automatically entered for selected drug.
• If selected drug had "_ _ / _ _ mL" concentration, drug
amount and diluent volume need to be entered.
• If selected drug is not weight-based, Not Used
displays in
PATIENT WEIGHT
-- Continued on Next Page --
2-32
Programming
- With Guardrails
Suite MX Protection (Continued)
®
Drug Library, the
®
) entry (if required)
BSA
soft key.
CONFIRM
field.
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
A
A
Guardrails Drugs
Peds Oncology
Amifostine
Azathioprine
Bleomycin
Cytarabine
Daunorbicin
>Select Drug
DRUG
PAGE
EXIT
DOWN
CALC
A
Guardrails Drug Setup
Peds Oncology
Methotrexate
_ _ _ mg/_ _ _mL was
selected.
Is this correct?
2
BSA
DOSING
mg/m
UNITS
>Press Yes or No
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
A-E
F-J
K-O
P-T
U-Z
Yes
No