Cardinal Health Alaris 8000 Series Directions For Use Manual page 68

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Preparing Syringe and Administration Set (
Priming (continued)
Pressure Sensing Disc Installed (Continued)
2.
Press
key.
OPTIONS
3.
Press Prime Set with Syringe soft key.
• If pressure sensing disc was not removed prior to
pressing Prime Set with Syringe soft key, a pressure
sensing disc removal prompt displays.
2-12
Getting Started
)
Syringe Module
(Continued)
SILENCE
OPTIONS
Alaris
Pump and Syringe Modules Section
FU
RATE (mL/h)
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
1
2
3
4
6
5
7
8
9
ENTER
.
0
CLEAR
CANCEL
A
Channel Options 1 of 2
Guardrails Drugs
Guardrails IV Fluids
Multidose
Pressure Limit - 550 mmHg
Prime Set with Syringe
>Select an Option or EXIT
PAGE
EXIT
DOWN
A A A
It is recommended that the
Pressure Sensing Disc be
removed during priming.
Press CONFIRM and
continue priming.
>Press CONFIRM
CONFIRM