Max Dose Limit
Module Location Enforcement
(Max Accumulated Dose Limit)
Near End of Infusion (NEOI)
NEOI Alert
Occlusion Pressure
Operating Modes
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
Features and Displays (Continued)
Optional configuration that limits total amount of drug allowed to
be delivered to patient in a defined period (
• This setting should be configured in data set before Drug
Library is developed. Once drugs are in Profile
Library, Max Accumulated Dose Limit cannot be changed.
• This optional setting applies to all drug setups within Profile
PCA Drug Library.
Tamper resistant security feature that ensures PCA Module is in a
tamper evident position. When enabled, PCA Module must be
located to direct right of PC Unit to allow programming an
infusion.
Allows an alert to be configured to sound anywhere between
5 – 25%
Alert Time can be set to occur when
Downstream occlusion alarm threshold can be set to low,
medium, or high.
Four operating modes are available:
PCA
• continuous infusion
PCA
• loading dose only
All programming of infusions in each of
using Drug Library as defined by hospital-established best-
practice.
volume remaining.
only
+ continuous infusion
,
or
hours).
1
2
4
Drug
PCA
of
remains.
5 – 25%
VTBI
modes are completed
4
General Information
3-45