Cardinal Health Alaris 8000 Series Directions For Use Manual page 90

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Primary Infusion
Intermittent Infusion (Continued)
6.
If an optional hospital-defined and editable starting value
for intermittent duration is not already entered, enter
duration or rate, as follows:
To enter duration, press
numeric data entry keys (rate value is calculated and
displayed).
• To enter rate, press
numeric data entry keys.
7.
Verify parameters are correct and press
• If programmed duration is outside Soft Limit for that
care area, a prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
If programmed duration is outside Hard Limit for that
care area, a prompt appears before programming can
continue. Infusion needs to be reprogrammed.
Pausing, Restarting, Restoring Infusion
Reference "Continuous Infusion" procedure.
Changing Rate or VTBI During Infusion
Reference "Continuous Infusion" procedure.
2-34
Programming
- With Guardrails
Suite MX Protection (Continued)
®
soft key and use
DURATION
soft key and use
RATE VOLUME
START
OR
soft key.
Alaris
Pump and Syringe Modules Section
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Setup
A
Methotrexate
PRIMARY INFUSION
4
RATE =
mL/h
ALL
VTBI
6:00
DURATION
hh:mm
1200
DOSE =
mg/m
2
[Conc]: 342 mg/mL
>Press START
DELAY
RATE
SETUP
START
OPTIONS
VOLUME
System Directions for Use
®

Advertisement

Table of Contents
loading