Primary Infusion
Intermittent Infusion (Continued)
6.
If an optional hospital-defined and editable starting value
for intermittent duration is not already entered, enter
duration or rate, as follows:
To enter duration, press
•
numeric data entry keys (rate value is calculated and
displayed).
• To enter rate, press
numeric data entry keys.
7.
Verify parameters are correct and press
• If programmed duration is outside Soft Limit for that
care area, a prompt appears before programming can
continue. If Yes soft key is pressed, programming
continues; if No soft key is pressed, infusion needs to
be reprogrammed.
If programmed duration is outside Hard Limit for that
•
care area, a prompt appears before programming can
continue. Infusion needs to be reprogrammed.
Pausing, Restarting, Restoring Infusion
Reference "Continuous Infusion" procedure.
Changing Rate or VTBI During Infusion
Reference "Continuous Infusion" procedure.
2-34
Programming
- With Guardrails
Suite MX Protection (Continued)
®
soft key and use
DURATION
soft key and use
RATE VOLUME
START
OR
soft key.
Alaris
Pump and Syringe Modules Section
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Setup
A
Methotrexate
PRIMARY INFUSION
4
RATE =
mL/h
ALL
VTBI
6:00
DURATION
hh:mm
1200
DOSE =
mg/m
2
[Conc]: 342 mg/mL
>Press START
DELAY
RATE
SETUP
START
OPTIONS
VOLUME
System Directions for Use
®