Anesthesia Mode (Continued)
• Bolus dose is automatically available for:
Guardrails
Drugs that have bolus dose limits defined
®
♦
generic drug calculation setup
♦
• Anesthesia Mode, alternating with other required prompts,
displays in prompt bar of Main Display.
• Callback audio for paused module is permanently silenced.
• Review of drug calculation setup page is omitted when
restoring a stopped drug calculation.
• Clinical Advisories are not displayed.
• Auto-ID Module is not available.
Enabling
1.
Press
OPTIONS
2.
Press Anesthesia Mode soft key.
3.
Press Enable soft key.
4.
Press
CONFIRM
Alaris
System Directions for Use
®
PC Unit Section
System Options (Continued)
key.
soft key.
System Options 1 of 2
Display Contrast
Patient ID
Time of Day
Power Down All Channels
Anesthesia Mode
>Select an Option or
EXIT
EXIT
System Options
Anesthesia Mode
Enable
Disable
Pump Module
Change
Air Detection =
75
microliters
>Select an Option or Press
CONFIRM
CONFIRM
General Setup and Operation
PAGE
DOWN
1-17