Setting Up Pca Dose + Continuous Infusion - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Setting Up PCA Dose + Continuous Infusion

1.
Perform steps in "Preparing Infusion".
2.
Press
PCA DOSE + CONTINUOUS
Mode screen.
3.
To enter
dose, press
PCA
numeric data entry keys.
4.
To enter lockout interval, press
key and use numeric data entry keys.
5.
To enter continuous dose, press
and use numeric data entry keys.
6.
To enter maximum limit, press
Yes soft key and use numeric data entry keys.
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
Infusion Modes (Continued)
soft key from Infusion
soft key and use
PCA DOSE
LOCKOUT INTERVAL
CONT DOSE
MAX LIMIT
soft
soft key,
soft key, press
A
Guardrails Drug Setup
Morphine
INFUSION MODES
PCA Dose only
Continuous Infusion
PCA Dose + Continuous
Loading Dose Only
>Select an Option
DRUG
SETUP
LIBRARY
PCA + Continuous
A
Morphine
1
PCA
mg
DOSE
LOCKOUT
_ _ _
minutes
INTERVAL
_ _ _
CONT
mg/h
DOSE
MAX
None
LIMIT
LOADING
None
DOSE
[Conc]: 1 mg/mL
>Enter Lockout Interval
CHANGE
MODE
PCA + Continuous
A
Morphine
1
PCA
mg
DOSE
10
LOCKOUT
minutes
INTERVAL
2
CONT
Yes
mg/h
DOSE
> > > >
MAX
No
LIMIT
LOADING
None
DOSE
[Conc]: 1 mg/mL
>Select Max Limit Option
CHANGE
MODE
Programming
3-17

Advertisement

Table of Contents
loading

Table of Contents