Battery Care And Maintenance - Cardinal Health Alaris 8000 Series Directions For Use Manual

Modular system intended for adult, pediatric and neonatal care in today’s growing professional healthcare environment
Table of Contents

Advertisement

Battery Type and Charging
The PC Unit is equipped with a
hydride battery. The battery is charging whenever the
instrument is plugged into an
expectancy of the battery is dependent on the amount of use,
the depth of discharge, and the state of the charge that is
maintained. Generally, the battery will have the longest life if
the instrument is plugged in and battery use is infrequent.
Frequent use of battery power and insufficient battery charge
cycles will significantly decrease the life of the battery.
The quality of the battery is also a significant factor in
determining battery life and runtime. The battery cannot be
repaired and should not be opened. Replace the battery with
the same type, size and voltage rating. Use of any other
brand may yield poor performance and is not recommended.
Batteries should be charged in a room with a temperature
between
-
°
50
80.6
F
maximize battery life.
• The PC Unit is shipped with the battery in a discharged
condition.
• Before the PC Unit is released for use, it should be plugged
into a hospital grade
at least
hours. This will ensure proper battery operation
8
when the Alaris
System is first set up for patient use.
®
• Whenever possible, leave the power cord connected to an
external
power source while operating the instrument.
AC
The battery capacity should be checked at least once every
months. Reference the Alaris
6
Manual for test and replacement procedures.
If the PC Unit is to be stored at temperatures in excess of
(
°
) for
or more months, the battery should be removed
30
C
1
and placed in an environment of
1-40
Troubleshooting and Maintenance

Battery Care and Maintenance

volt,
12
4000
receptacle. The life
AC
(
-
°
), to minimize charge time and
10
27
C
Battery Charge
outlet and the battery charged for
AC
Battery Care
System Technical Service
®
-
°
50
86
mAh nickel metal
86
(
-
°
).
F
10
30
C
°
F
Alaris
System Directions for Use
®
PC Unit Section