8210 Series (Continued)
When saturation is calculated from a blood gas partial
pressure of oxygen (
the
p
measurement of the
S
O
2
occurs because the calculated saturation was not
appropriately corrected for the effects of variables that shift the
relationship between
pressure of carbon dioxide (
hemoglobin. To compare functional saturation measurements
to those from an instrument that measures fractional
saturation, fractional measurements must be converted as
follows:
Oxyhemoglobin Dissociation Curve
pH
­
100
Temperature
¯
PCO
¯
2,3-DPG
¯
Fetal Hb
­
50
0
4-32
General Information
Principle of Operation (Continued)
), the calculated value may differ from
PO
2
p
Module, this usually
S
O
2
and p
, temperature, the partial
PO
H
2
),
PCO
2,3-DPG
2
2
¬
®
pH
­
Temperature
¯
PCO
¯
2,3-DPG
¯
50
PO (mmHg)
2
, and fetal
2
100
(8210)
Alaris
System Directions for Use
®
SpO
Modules Section
2