Messages (Continued)
Message
NEOI (Near End of Infusion)
Panel Locked
Panel Unlocked
Pause
PCA Complete
PCA Not In Secure Location
Syringe Not Recognized
NOTE:
The device will re-alarm if the Dose Request Cord is pressed
again while in a Max Limit state.
Alaris
System Directions for Use
®
Patient Controlled Analgesia Module Section
Alarms, Errors, Messages (Continued)
Syringe almost empty.
Tamper Resist feature is active and
a key was pressed.
Tamper Resist feature deactivated.
Pause control pressed; infusion
stopped.
Current
Lockout interval begins. Channel
running in continuous dose if
programmed.
PCA Module is not in preferable
location to allow locking to PC Unit.
Device is not in a tamper evident
position.
Installed syringe of unknown type
and size.
Meaning
dose complete.
PCA
Response
None. This is a timed event that
can be set. To set or change this
option, reference "General
Information", "Configurable
Settings".
If appropriate, deactivate Tamper
Resist feature using Tamper Resist
Control on back of PC Unit.
None.
To resume infusion, press
key, or press
RESTART
key and then
SELECT
key.
None.
Detach PCA Module from current
position and reattach to immediate
right of PC Unit.
Select and confirm correct syringe
type and size, and then press
; or use a syringe type
CONFIRM
and size that system can
automatically and correctly identify.
Troubleshooting and Maintenance