Primary Infusion
Bolus Dose (Continued)
7.
Verify parameters are correct and press
• If programmed bolus dose and/or bolus dose duration
is outside Soft Limit for that care area, a prompt
appears before programming can continue. If Yes soft
key is pressed, programming continues; if No soft key
is pressed, infusion needs to be reprogrammed.
If programmed bolus dose and/or bolus dose duration
•
is outside Hard Limit for that care area, a prompt
appears before programming can continue. Infusion
needs to be reprogrammed.
If a bolus dose outside of Soft Limits has been entered
•
and verified as correct, Message Display also shows
either "
" for a low dose or "
LLL
Stopping Bolus Dose
The display examples in this procedure represent stopping a
bolus dose which was programmed using the Drug Library.
Even where the displays are different when stopping a bolus
dose which was programmed using a non-library drug, the
procedure is the same.
1.
Press
CHANNEL SELECT
2.
Press
STOP BOLUS
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
- With Guardrails
®
START
" for a high dose.
↑↑↑
key.
soft key.
Suite MX Protection (Continued)
soft key.
FU
FU
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
SELECT
PAUSE
PAUSE
CHANNEL
OFF
CHANNEL
OFF
RESTART
RESTART
Guardrails Drug Library
Guardrails Drug Setup
A A A
A A A
Dopamine
Alteplase
BOLUS DOSE
BOLUS DOSE
5
0.9
mcg/kg
DOSE
DOSE
mg/kg
71
PATIENT
PATIENT
70
kg
kg
WEIGHT
WEIGHT
(420 mL/h)
<1
1
Rapid
DURATION
DURATION
min
min
Bolus
INFUSE AT: 0.09 mg/kg/min
BOLUS VTBI = 50.0 mL
[Conc]: 1 mg/mL
[Conc]: 800 mcg/mL
>Press Start to Begin
>Press START to Begin
Infusing Bolus Dose
Infusing Bolus Dose
CONT-
CONT-
SETUP
SETUP
START
START
PAUSE
PAUSE
INUOUS
INUOUS
A A A
Guardrails Drug Setup
Alteplase
BOLUS DOSE
Dose
0.9
Remaining:
mg/kg
70
Pt. Weight:
kg
1
Time Left:
min
TOTAL DOSE = 6.3 mg
BOLUS VTBI = 6.3 mL
>Press START to Continue
Infusing Bolus Dose
STOP
PAUSE
START
BOLUS
Programming
2-29