While Infusion is in Progress (Continued)
3.
Press Patient ID soft key.
• Patient ID Entry screen appears.
• Green
READY LED
ready to scan.
4.
To scan bar code on patient identification band, press
scan trigger on hand-held scanner.
• If scan is successful, an audible tone sounds and
patient
displays on Main Display.
ID
NOTE:
Patient
may be entered manually using the PC Unit keypad
ID
(reference PC Unit Section of this
Authorized User Mode is a feature that combines the PC Unit
tamper resist feature with the Auto-ID application. This
feature is designed to ensure that only clinicians with a bar
code on their
badge can program the Alaris
ID
When this feature is enabled, the PC Unit automatically
enables tamper resist mode upon power on and
after programming is completed. To unlock the keypad, the
user must scan their
To power on the PC Unit with Authorized User Mode enabled:
1.
Power on system and associate patient
"Patient Identification" procedure).
• Upon successful entry of patient
automatically enables tamper resist feature.
2.
To disable tamper resist, press
badge.
Alaris
System Directions for Use
®
Auto-ID Module Section
Patient Identification (Continued)
illuminates, indicating system is
).
DFU

Authorized User Mode

badge.
ID
ID
SCAN
System.
®
minutes
5
(reference
ID
, PC Unit
key and scan
ID
CAUTIONS
CLASS 1 LED PRODUCT
stare into the beam or allow
beam to strike patient's face.
• Always verify that information
displayed on the PC Unit
matches scanned data.
Programming
: Do not
6-5