Secondary Infusion
This mode is designed to support automatic secondary
infusions ("piggybacking") in the same instrument. A
secondary infusion can be programmed as a "Basic Infusion"
or "Drug Library Infusion". When the secondary
zero, an audio tone will sound (if enabled) indicating
completion of the secondary infusion. The primary infusion
resumes automatically.
When the instrument is programmed and delivering in the
secondary mode, the primary infusion is temporarily stopped
and fluid is drawn from the secondary container. Delivery
from the primary container resumes when the fluid level in the
secondary line is level with the fluid in the primary container.
1.
Open secondary administration set package, remove set
and close clamp.
2.
Insert administration set spike into prepared fluid
container and hang secondary container, following
accepted hospital/facility procedure.
3.
Fill drip chamber to
4.
Open secondary administration set clamp and prime set.
Close clamp.
5.
Attach secondary administration set to upper injection site
on primary set.
Alaris
System Directions for Use
®
Pump and Syringe Modules Section
- With Guardrails
(
Pump Module
Introduction
Setup
/
full.
2
3
Suite MX Protection
®
)
WARNINGS
• Secondary applications require
• The secondary solution
reaches
VTBI
• The secondary
• The clamp on the secondary
• The secondary administration set
FU
CHANNEL
SELECT
PAUSE
CHANNEL
OFF
RESTART
the use of a check valve set on
the primary
line.
IV
container must be higher than
the primary solution container.
VTBI
require consideration of such
variables as factory overfill,
medication additions, etc.
Underestimating the volume will
cause the remaining secondary
solution to be infused at the
primary rate; overestimating will
result in the primary solution
being infused at the secondary
rate. Multiple doses from a
single container are not possible.
administration set must be
opened. If the clamp is not
opened, the fluid will be
delivered from the primary
container.
must be primed prior to
beginning the secondary
infusion.
Programming
settings
2-39